Setup mode
OPERATION MANUAL Chapter 4
CC® Liability for errors and misprints excluded. V1.3.0en/06.07.15//14.06
62
4.2.11 Sub-category: “Display parameters
In this function, the set parameters are displayed. Depending on the previous setting, these will be
the “Automatic control parameters” or the “Manual control parameters”.
PROCEDURE
Go to the “Categories Menu”.
Tap on the category “Temperature Control”.
Tap on the category “TAC/Manual”.
Tap on the sub-category “Display parameters”.
Tap on “OK” after you have read/checked the parameters.
4.2.12 Einfrierschutz für Temperiergeräte (nicht gültig für CC-520w, CC-525w, CC-
906w und Kältebäder im Dauerbetrieb)
Non-compliance with the compatibility between the thermal fluid and your temperature control
unit
PROPERTY DAMAGE
Water with a temperature below 0 °C and without an anti-freeze must not be used as a ther-
mal fluid (risk of freezing and potential destruction of the evaporator).
Water below a temperature of 5 °C must not be used without an anti-freeze as a thermal fluid
(risk of freezing and potential destruction of the evaporator).
Observe the rating of your temperature control unit in accordance with DIN 12876.
Ensure the following materials are resistant with respect to the thermofluid: Stainless steel
1.4301/ 1.4401 (V2A), copper, nickel, FKM, red bronze/brass, silver solder and plastic.
The maximum viscosity of the thermal fluid must not exceed 50 mm²/s at the lowest working
temperature!
The maximum density of the thermal fluid may not exceed 1 kg/dm³!
Operation without freeze protection activated
COMPLETE DESTRUCTION OF THE TEMPERATURE CONTROL UNIT
When using water as a thermal fluid, freeze protection must be activated in the category “pro-
tective devices”.
Temperature control units with freeze protection option can use water as the only thermal fluid.
Possible danger: Freezing of the heat exchanger at < 5 °C. An ice crystal (green flashing) indicates the
activation of the freeze protection on the touch display. To find out whether your temperature
control unit is equipped with a freeze protection, refer to the datasheet from page 83 in section
»Annex«.
PROCEDURE
Go to the “Categories Menu”.
Tap on the category “Protection Options”.
Tap on the category “Freeze Protection”.
Either tap on the dialog entry “off” or “on” to turn the freeze protection on or off.
Confirm your choice by tapping on “OK”.