Front Panel
LASer:LIMit:Il
W
Remote
The LASer:LIMit:Il command sets the LASER current limit value for the
100
rnA
range on the LDC-3712
and the 1000
rnA
range on the LDC-3742B.
(Not
~pplicable for LDC-3722B).
SYNTAX
DIAGRAM
POINTS OF
INTEREST
Front Panel
W
Remote
An
<nrf
value> which represents the LASER
limit
current, in
mA.
The current limit is in effect in all modes of operation of the laser output.
In local operation, the limit current is entered by selecting LIM
I
parameter, pressing and
holding in the (PARAMETER) SET switch, adjusting the ADJUST knob until the desired
value
is
displayed, and then releasing the
SET
switch.
"LAS:LIM:Il
80"
-action: the LASER current limit is set to
80
rnA.
":Laser:Limit:Il 60" -action: the LASER current limit is set to
60
mA.
The
LASer:LMit:Il? query returns the value of the LASER current limit for the
100
rnA
range on the
LDC-3712 or the 1000
mA
range on the LDC-3742B. (Not applicable for LDC-3722B).
SYNTAX
DIAGRAM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ILX Lightwave LDC-3700 Series and is the answer not in the manual?

ILX Lightwave LDC-3700 Series Specifications

General IconGeneral
BrandILX Lightwave
ModelLDC-3700 Series
CategoryController
LanguageEnglish

Related product manuals