1 - 6
Transpector XPR 3+ Operating Manual
1.9.2 Weight
Transpector XPR 3+ electronics module weighs 1.62 kg (3.58 lbs).
1.9.3 Mounting Requirements
The sensor is mounted to a high-vacuum chamber with a standard 2.75 in.
(69.9 mm) O.D. ConFlat flange.
The electronics module attaches to and is supported by the sensor.
Transpector XPR 3+ can be mounted in any position. See section 1.16.5, Sensor
Installation, on page 1-19 for information on installing Transpector XPR 3+ system.
1.9.4 Ventilation Requirements
At least 25.4 mm (1 in.) of open space around the Transpector XPR 3+ electronics
module must be maintained for proper ventilation.
1.9.5 Maintenance Access
Easy access to Transpector XPR 3+ should be maintained for installation and
maintenance activities.
1.10 Electrical Power Requirements
Transpector XPR 3+ must be connected to a source of power as specified in the
following sections.
1.10.1 Required Supply Voltage
20 to 30 V(dc), 24 V(dc) typical
1.10.2 Current Rating
1.25 A maximum
1.10.3 Electrical Connection
Latching, 4-pin DIN connector, internally isolated from system ground.
See Figure 1-3.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inficon Transpector XPR 3+ and is the answer not in the manual?

Inficon Transpector XPR 3+ Specifications

General IconGeneral
BrandInficon
ModelTranspector XPR 3+
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals