3 - 3
Transpector XPR 3+ Operating Manual
3.4  Sensor
The Transpector XPR 3+ sensor (see Figure 3-2) analyzes gases by ionizing some 
of the gas molecules (in the ion source), separating the ions by mass (in the mass 
filter), and measuring the quantity of ions at each mass (in the detector). The 
masses, unique for each substance, identify the gas molecules from which the ions 
were created. The magnitudes of these signals are used to determine the partial 
pressures (amounts) of the respective gases.
The Sensor consists of three main parts: 
 ion source (ionizer)
 quadrupole mass filter
 ion detector
All of these parts are mounted on an electrical feed-through flange, which is bolted 
to the vacuum chamber where the gas analysis measurements are made.
The sensor works only in a high-vacuum environment because the ions, once 
created, must not collide with other gas molecules as they move through the 
sensor; otherwise, they might not be detected. The miniature design of 
Transpector XPR 3+ allows it to operate at pressures higher than those necessary 
for traditional RGA sensors.