6 - 1
Transpector XPR 3+ Operating Manual
Chapter 6
Transpector XPR 3+ Low Pressure EM
6.1  Introduction
This chapter provides information concerning the Transpector XPR 3+ Low 
Pressure EM option. Information concerning the software is located in the 
FabGuard Explorer Operating Manual or the FabGuard help files, included with 
your Transpector XPR 3+ Gas Analysis System.
6.2  Transpector XPR 3+ Filament Caution
Sampling atmosphere or oxygen above 1x10
-4
 Torr using Transpector XPR 3+ is 
not recommended. To obtain maximum useful life from the Transpector XPR 3+ 
filaments observe the following caution.
Attempting to turn on the emission above 10 mTorr 
(20 mTorr with Pirani Interlock option) may result in 
premature failure of filaments.
Utilize the Pirani Interlock option or other measures to 
operate the Transpector XPR 3+ within a safe pressure range.
Failure to observe this caution will result in premature failure of filaments. Filament 
replacement is not covered by warranty under these conditions. 
Please read Chapter 3, How The Instrument Works, before using the Transpector 
XPR 3+ filament and follow the recommendations in the Transpector XPR 3+ 
Operating Manual, Chapter 6, Best Known Methods.
6.3  Electron Multiplier Caution
Operating the Electron Multiplier at a total pressure 
greater than 1x10
-4
 Torr, or with an output current greater 
than 5x10
-7
 Amps, will damage the Electron Multiplier and 
reduce its lifetime.