3 - 10
Transpector XPR 3+ Operating Manual
When the MCP is grounded, the ions exiting the quadrupole through the exit lens 
are collected on the Faraday Cup. The resulting current is conducted through the 
signal output to the detection amplifier. When -1275 V is applied to the front of the 
MCP, and between -500 and -50 V is applied to the back of the MCP, the ions 
impinge on the front side of the MCP. The resulting electron current is collected by 
the same Faraday electrode.
The front of the MCP is fixed at -1275 V in the EM mode for two reasons. First, the 
ion beam exiting the quadrupole can be strongly divergent, -1275 V ensures that 
the entire ion beam is deflected into the MCP. Second, if the ion’s kinetic energy as 
it strikes the entrance of the EM is too low, severe mass discrimination effects can 
occur. The -1275 V avoids both issues.
Use the minimum MCP voltage required to obtain the necessary peak amplitudes 
and/or signal-to-noise ratio—a gain of 300 is recommended. Operating at higher 
voltages than necessary will result in premature aging of the Electron Multiplier, 
requiring early replacement. As the MCP ages, the voltage needed to get a specific 
EM gain will increase.
3.5  Scanning Characteristics
As described above, the quadrupole acts as a mass filter for a mixed beam of 
ions, rejecting those of both high and low mass, while passing those of an 
intermediate mass. The selectivity of the mass filter is expressed in terms of 
resolution, R, which is numerically given by the ratio of the center mass, M, to the 
width, M (both in AMU), of the pass band. Since the number of the ions passed 
by the filter falls off gradually as the edge of the pass band is approached, the 
width is defined at the point where the ion current falls to some specified fraction 
(usually 1/2 or 1/10) of the maximum value. The width of the pass band is 
determined by the DC-to-RF ratio.
While the quadrupole drive circuits can be designed so that R varies in any desired 
manner with M, it is usually most convenient to keep M constant at a value, which 
ensures adequate separation of masses that are 1 AMU apart. This mode of 
scanning is called Constant M. As a result, R is proportional to M, and therefore 
the efficiency with which ions of mass M are transmitted through the quadrupole 
decreases with M. Thus, the sensitivity of the sensor decreases as M increases.
3.6  The Zero Blast
When the mass filter is tuned to very low masses, the RF and DC voltages applied 
to the rods approach zero. The quadrupole then ceases to act as a filter, and a 
large current of unseparated ions is detected. This current is called the “zero blast."