Multitron - Operating Manual
Technical Data and Specifications
Page 148 of 158 20 August 2020
12.5.3 Direct Steam Humidification
Casing dimensions
Description Valu
e
Unit
Height
340
mm
Depth
80
mm
Width
200
mm
Various
Description
Value
Unit
Additional mass
3
kg
Water consumption
1
8.3
g/h
Temperature range for use
2
Up to 40
°C
Heating time of the unit
5
min
1
) At an ambient temperature of up to 20 °C and at a temperature of 37 °C and
relative humidity of 85 % in the incubation chamber.
2
) Temperature in the interior at which functioning can be guaranteed
Target and actual humidity values
Description
Value
Unit
Setting range
20.0 to 85.0
% rH
Increment
1.0
% rH
Control precision
± 3.0
%r H
Values that can be reached
(at 37 °C in the interior and 20 °C room
temperature)
85.0
% rH
Electrical data
Description
Value
Unit
Voltage
230
V
Max. power consumption
125
W
Max. current consumption
0,4
A
Bar heater (in door)
24
V
15
W
Door seal heating
24
V
17
W
Floor heating
24
V
17
W

Table of Contents