Multitron - Operating Manual
Technical Data and Specifications
20 August 2020 Page 149 of 158
Specifications for the connection / water quality
Description
Value
Unit
Connection (hose Ø)
6 to 7
mm
Input pressure
0 to 0.3
bar
Water hardness
< 0.01, CaCO
3
equivalent
mmol/L
Water conductivity
min. 5 / max. 20
μS/cm
Dissolved solids
< 10
mg/L
ATTENTION
Using tap water can quickly lead to limescale in the vaporiser of
the humidification unit, which would affect correct functioning.
Ultra pure water (WFI, water for injection purposes) must not be
used either as this is highly corrosive and thus damages the device.
Adding detergents, disinfectants or similar chemicals to the water
for the direct steam humidification can damage the device. Only
use water without additions, in accordance to the specification.
Specifications for the humidity sensor
Description
Value
Unit
Sensor type EE071 HCT01-00D
Measurement range
20 to 90
%
Response time
< 300
ms
Temperature range for use -
40 to 80
°C
12.5.4 Pressure Elimination Unit for Direct Steam Humidification
Dimensions
Description
Value
Unit
Depth
185
mm
Height
337
mm
INFORMATION
For detailed information on the external dimensions, refer to the
dimension drawings (see chapter 4.3.4 "Dimensions", page 47).