Clock/battery errors
Clock or battery errors are displayed together with
the "EEE" symbol followed by "bAt" and a status number.
For a more detailed description please refer to "Baterry
control function" and "Replacing the battery".
Fig. Clock / battery error
number 330 (clock does
not run and has to be set.
Major errors
When a major error occurs, the device must be sent
to the manufacturer's service center for inspection
and adjustment.
Errors Error description
EEE
910
Error while reading the calibration.
Errors Error description
0x01 EEPROM does not respond.
0x02 Address overrange.
0x04 Checksum error.
0x08 Error in the internal I2C bus.
Example, error message 911:
The error number consists of major error 910 and internal
error cause 0x01 .
In this example an error
occurred while reading
the calibration data from
EEPROM. The device
has to be returned
to the manufacturer
for inspection and
adjustment.
Internal causes:
The UMG 96RM-E sometimes determines the cause
of a major internal error with the following error code.