Chapter 1. Setting Up the RF ValProbe
10 RF ValProbe® User’s Manual
1.2.3 Required Equipment
When verifying sensor calibration, you must consider the measurement uncertainty of the
equipment you are using in combination with the uncertainty of the equipment used by
Kaye to calibrate the Loggers at the factory.
For example, if you verify calibration for a Logger's humidity sensor using a calibration
chamber with a measurement uncertainty of ±1.5% RH, and the Kaye chamber has a
measurement uncertainty of ±1.0% RH, you may obtain a deviation greater than the 2.0%
RH specified by the factory as the accuracy of the sensor.
The Certificate of Calibration that was shipped with each Logger provides detailed
information about the equipment used by the factory to calibrate the Loggers, including
the measurement uncertainty. (Auxiliary inputs are shipped with additional certification
documents.) The equipment required to accurately verify calibration for each sensor type
is listed on the next page.