Appendix A. Specifications
RF ValProbe® Users Manual 51
Appendix A. Specifications
A.1 RF ValProbe System Specifications
A.1.1 RF System Operating Frequency
2.4 GHz
A.1.2 RF Certification
USA, Canada, European Union, China, India, Brazil, Argentina, Japan, Australia, Israel,
Indonesia, Singapore, South Korea, South Africa. For other countries, contact factory.
A.1.3 Number of Loggers
Up to a maximum of 40 Loggers per Base Station.
A.1.4 Transmission
Maximum antenna to antenna distance in open space: 300 feet
A.1.5 PC Update Rate
10 seconds maximum
A.1.6 Maximum Network Form Time
For 8 Loggers evenly distributed over 5 cubic ft.: typical, 2 minutes
For 12 Loggers in a 30 ft radius single plane, typical, 4 minutes
A.1.7 Data Retention
Loggers retain up to 10,000 data samples.