EasyManua.ls Logo

Kaye RF ValProbe - Logger Installation Guidelines; Whats Next

Kaye RF ValProbe
103 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 1. Setting Up the RF ValProbe
16 RF ValProbe® User’s Manual
1.3.4 Logger Installation Guidelines
You can now position the Loggers. Each Logger must be no more than 300 ft. distant from
the Base Station and the other Loggers. Kaye offers several recommendations for logger
placement:
Install the Loggers at least one to two feet above the ground or floor. For better RF
transmission, place them as high as possible. If possible, raise or lower the Base
Station and Loggers above or below walls or any separation.
Point the Logger antenna upwards for best results.
Do not position the Loggers directly above or below each other. Stagger their
positions for better transmission.
If Loggers are placed at different heights, make sure that they are within antenna range
(within 300 ft. from the Base Station or another logger). The RF signal is transmitted
in an arc, with maximum signal strength occurring in the area 45° above and below
the tip of the antenna.
Be aware that metal surroundings can interfere with RF transmission; the Logger may
transmit, but the signal will be weaker. Shorten the transmission distance accordingly.
Do not install Loggers next to a cordless phone base or other 2.4 GHz transmission
device.
Do not install Loggers on a vibrating surface.
Do not place Loggers where the temperature is outside their rated operating range.
IMPORTANT: If you are using Loggers with internal sensors, be sure the vents on the side
remain uncovered. If the vents are covered, the sensors cannot provide
correct readings.
1.3.5 What’s Next?
Once you have connected the Base Station and positioned the Loggers, proceed to Chapter
2 to use the RF ValProbe software and begin taking measurements.

Table of Contents