Appendix B. Using the OPC Server and Client
RF ValProbe® Users Manual 69
B.5.2 Optimizing Network Performance (cont.)
a. To find the current frame size, click the Get button to the right of the Frame Size
box.
b. To set a particular size as the frame size, click on the size from the drop-down list
and click the
Set button. A message appears, telling you that the frame size has been
successfully set.
The Network ID number for each sensor (from 0 to 9, or A to F) is set on a small
rotating dial at the back of the sensor. (See page 15.)
a. To obtain the established Network ID, click the
Get button to the right of the
Network ID box.
b. To set another number as the ID, enter the number in the text box and click the
Set
button. A message appears, telling you that the Network ID has been successfully
established.
Click
OK to return to the Device window. You have now set up your RF ValProbe and
sensors for viewing on the Sensors window and the OPC Client.
Note: Once you set or change parameters in the Advanced window, the server will issue
a command to reset the system. Sensors will disconnect from and rejoin the
network, creating an interval before data transmission resumes.