SCPI Tables 11-5
Table 11-4 (cont.)
SENSe command summary
Command Description
Default
parameter
SCPI
:TLEVel?
:TLEVel
[:AMP] <NRf>
[:AMP]?
:MILLiamp <NRf>
:MILLiamp?
:LINTegration
:TIME <NRf>
:TIME?
:AUTO
:TLEVel <NRf>
:TLEVel?
:TLEVel
[:AMP] <NRf>
[:AMP]?
:MILLiamp <NRf>
:MILLiamp?
:TEDGe <name>
:TEDGe?
:TimeOUT <NRf>
:TimeOUT?
:SEARch <b>
:SEARch?
:FAST <b>
:FAST?
Model 2303 and 2303B - query trigger level.
Model 2303-PJ:
Set trigger level for 5A range: 0 to 5 (amps).
Query trigger level for 5A range.
Set trigger level for 500mA range: 0 to 0.5 (amps).
Query trigger level for 500mA range.
Path to configure long integration measurements:
Set integration time in seconds: X to 60
(where X is 0.850 for 60Hz or 0.840 for 50Hz).
Query integration time.
Power supply sets integration time.
Models 2303 and 2303B - Set trigger level: 0 to 5A.
Models 2303 and 2303B - Query trigger level.
Model 2303-PJ:
Set trigger level for 5A range: 0 to 5 (amps).
Query trigger level for 5A range.
Set trigger level for 500mA range: 0 to 0.5 (amps).
Query trigger level for 500mA range.
Select trigger edge to initiate the measurement:
RISING, FALLING or NEITHER.
Query trigger edge.
Specify length of timeout: 1 to 63 (seconds).
Query timeout.
Enable or disable pulse search.
Query state of pulse search.
Enable or disable long integration fast readings mode.
Query long integration fast readings mode.
0
0
1
0
0
0
RISING
16
ON
OFF