C-4 Emulation Commands
VOUT?
Trigger and return an output voltage reading
Description
When this command is sent, the power supply is triggered to measure the out-
put voltage. After the power supply is addressed to talk, the single voltage
reading (in x.xxx format) is sent to the computer. This changes the function
of the supply to be set for voltage.
IOUT?
Trigger and return an output current reading
Description
When this command is sent, the power supply is triggered to measure the out-
put current. Before the measurement is performed, the 5A measurement
range is selected (AUTO range disabled).
After the power supply is addressed to talk, the single current reading (in
y.yyyy format) is sent to the computer. This changes the function of the sup-
ply to be set for current.
ID?
Identification
Description
This query responds with the model number of the instrument (KI2303,
KI2303B, or KI2303-PJ).