14-10 SCPI Reference Tables Model 6220/6221 Reference Manual
Return to Section 14 topics
:CABort <b> Enable to abort test on compliance: ON or OFF OFF
:CABort? Query compliance sweep abort state.
:CSWitch <b> Enable cold switching mode: ON or OFF. OFF
:CSWitch? Query cold switching state.
:ARM Arm Delta.
:ARM? Query arm. 1 = Delta armed, 0 = not armed.
:PDELta Pulse Delta operation: 5
:NVPResent? Query connection to 2182A: 1 = yes, 0 = no.
:HIGH <NRf> Set high pulse value (amps): -105e-3 to 105e-3. 1e3
:HIGH? Query high pulse value.
:LOW <NRf> Set low pulse value (amps): -105e-3 to 105e-3. 0
:LOW? Query low pulse value.
:WIDTh <NRf> Set pulse width (seconds): 50e-6 to 12e-3. 110e-6
:WIDTh? Query pulse width.
:SDELay <NRf> Set source delay (seconds): 16e-6 to 11.966e-3. 16e-6
:SDELay? Query source delay.
:COUNt <NRf> Set number of Pulse Delta readings to perform:
1 to 65636 or INFinity.
INF
:COUNt? Query Pulse Delta count.
:RANGing <NRf> Select fixed pulse source range: BEST or FIXed. BEST
:RANGing? Query fixed pulse source range.
:INTerval <NRf> Set interval (in PLCs) for each pulse cycle: 5 to
999999.
5
:INTerval? Query interval.
:SWEep Pulse Delta Sweep output:
[:STATe] <b> Enable or disable Sweep output mode. OFF
[:STATe]? Query state of Sweep output mode.
:LMEasure <NRf> Set number of low measurement per cycle: 1 or 2. 2
:LMEasure? Query number of low measurements.
:ARM Arm Pulse Delta.
:ARM? Query arm. 1 = Pulse Delta armed, 0 = not
armed.
Table 14-6 (cont.)
Source command summary
Command Description Default Sec SCPI