EasyManuals Logo

Keithley 6220 User Manual

Keithley 6220
454 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #395 background imageLoading...
Page #395 background image
Model 6220/6221 Reference Manual Calibration 17-19
Return to Section 17 topics
Step 4: Calibrate guard circuit
1. Connect the digital multimeter to the Model 6220/6221 GUARD and LO
jacks as shown in Figure 17-4. Be sure to connect the HI and LO terminals
of the Model 6220/6221 OUTPUT jack together as shown. Select the DMM
DC voltage function with auto range enabled.
2. Send these commands in order to calibrate the guard circuit using the
<DMM_Reading> parameter (actual digital multimeter reading) for each
step:
Step 5: Send dates, save calibration, and lock output calibration
Send these command to send dates, save calibration constants and dates, and
then finally lock out calibration:
SOUR:CURR:RANG 2e-4 ‘ Set current range to 200μA.
SOUR:CURR 2e-5 ‘ Set current source to 20μA.
OUTP ON ‘ Turn on output.
CAL:PROT:GUAR:STEP1:INIT ‘ Set up for step 1 guard cal.
CAL:PROT:GUAR:STEP1 <DMM_Reading> ‘ Cal step 1 with reading.
CAL:PROT:GUAR:STEP2:INIT ‘ Set up for step 2 guard cal.
CAL:PROT:GUAR:STEP2 <DMM_Reading> ‘ Cal step 2 with reading.
OUTP OFF ‘ Turn off output.
CAL:PROT:DATE 2004,05,15 ‘ Send cal date (5/15/2004).
CAL:PROT:NDUE 2005,05,15 ‘ Send cal due date (5/15/2005).
CAL:PROT:SAVE ‘ Save cal constants and dates.
CAL:PROT:LOCK ‘ Lock out calibration.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6220 and is the answer not in the manual?

Keithley 6220 Specifications

General IconGeneral
BrandKeithley
Model6220
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals