EasyManua.ls Logo

Keithley 6514

Keithley 6514
334 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Triggering 9-3
The only difference between front panel operation (Figure 9-1) and remote operation (Figure
9-2) is within the idle state of the instrument. Nomenclature in Figure 9-1 relates to the various
names used for configuration menu items, while Figure 9-2 provides the SCPI commands to
control operation.
Arm-In
Event
Arm Event
Detector
Trigger Event
Detector
Trigger Delay
MEASURE
Action
No
Yes
Idle
Another
Arm
?
Another
Trigger
?
Yes
No
Arm
Layer
Trigger
Layer
= GPIB Default
= Output Trigger
Trigger-In
Source
0.0 sec
No
See Note
INITiate
?
Yes
SOURce
ARM
:DIRection
ACCeptor
ARM:SOURce IMMediate
ARM:SOURce BUS
ARM:SOURce TIMer
ARM:SOURce MANual
ARM:SOURce TLINk
ARM:SOURce NSTest
ARM:SOURce PSTest
ARM:SOURce BSTest
TRIGger
:DIRection
SOURce
ACCeptor
TRIGger:DELay <n>
TRIGger:OUTPut
SENSe | NONE
TRIGger:COUNt <n>
ARM:OUTPut
TRIGger|NONE
ARM:COUNt <n>
TRIGger:SOURce IMMediate
TRIGger:SOURce TLINK
Note:
The following commands place the Model 6514 into
idle: ABORt, *RST, SYSTem:PRESet, *RCL <NRf>,
DCL, and SDC.
TRIGger:DELay:AUTO <b>
1
1
Figure 9-2
Trigger model —
remote operation

Table of Contents

Other manuals for Keithley 6514

Related product manuals