EasyManuals Logo
Home>Keithley>Measuring Instruments>6514

Keithley 6514 User Manual

Keithley 6514
334 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #272 background imageLoading...
Page #272 background image
Calibration 19-19
4. Enter the present calibration code on the display. (Factory default: 006514.) Use the up
and down RANGE keys to select the letter or number, and use the left and right arrow
keys to choose the position. Press ENTER to complete the process, and the unit will
display:
NEW CODE? Y/N
5. Select Y, then press ENTER. The unit will prompt for the new code:
CODE? 000000
6. Enter the new code, then press ENTER.
7. Using the LOCK selection in the calibration menu, lock out calibration after changing
the code.
Resetting the calibration code
If you forget the calibration code, you can unlock calibration by shorting together the CAL
pads, which are located on the display circuit board inside the unit. Doing so will also reset the
code to the factory default (006514).
Displaying calibration dates
To display calibration dates at any time:
1. From normal display, press SHIFT then CAL. The unit will display the following:
CAL: RUN
2. Use either RANGE key to select CAL: DATES, then press ENTER. Model 6514 will dis-
play the last calibration date, for example:
DATE: 06/15/98
3. Press ENTER to view the calibration due date, for example:
NDUE: 06/15/99
4. Press EXIT to return to normal display.

Table of Contents

Other manuals for Keithley 6514

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6514 and is the answer not in the manual?

Keithley 6514 Specifications

General IconGeneral
BrandKeithley
Model6514
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals