2-18 Measurement Concepts
SCPI programming
Table 2-5 lists SCPI commands used for input bias current and offset voltage calibration.
SCPI command input bias current calibration
1. Connect a triax shielding cap to the rear panel INPUT jack. (Use a Keithley CAP-31 or
equivalent.)
2. Send the following command to perform input bias current calibration:
:CAL:UNPR:IOFF
3. Allow the Model 6514 to complete the calibration process.
SCPI command offset voltage calibration
1. Connect a triax short to the rear panel INPUT jack. (Use the supplied Model 237-ALG-2
triax cable or equivalent with red and black alligator clips connected together.)
2. Send the following command to perform offset voltage calibration:
:CAL:UNPR:VOFF
3. Allow the Model 6514 to complete the calibration process.
Table 2-5
SCPI commands — input bias current and offset voltage calibration
Commands Description
:CALibration:UNPRotected:IOFFset
:CALibration:UNPRotected:VOFFset
Input bias current calibration.
Offset voltage calibration.

Table of Contents

Other manuals for Keithley 6514

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6514 and is the answer not in the manual?

Keithley 6514 Specifications

General IconGeneral
BrandKeithley
Model6514
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals