Measurement Concepts 2-3
SCPI programming
Programming example
The following command sequence will perform one zero corrected amps measurement:
SYST:AZER OFF ‘ Disable autozero.
SYST:AZER ON ‘ Enable autozero.
Connection fundamentals
The following provides important fundamental information on input connections to Model
6514. Typical connection drawings are included with the various measurement procedures pro-
vided in subsequent sections of this manual.
Input connector
The rear panel INPUT connector is a 3-lug female triax connector that will mate to a cable
terminated with a male triax connector.
Table 2-2
SCPI commands — autozero
Command Description Default
SYSTem SYSTem Subsystem:
:AZERo
[:STATe] <b> Enable or disable autozero. ON

Table of Contents