19-12 Calibration
2. Select the amps function on Model 6514.
3. Select Model 6514 20pA range.
4. Press SHIFT then CAL, then press ENTER at the CAL: RUN prompt. The unit will
prompt for the zero calibration point:
20PA ZERO
5. Allow the settling time listed in Table 19-5, then press ENTER.
6. The instrument will prompt for the positive full-scale calibration point:
+20PA CAL
7. Connect the voltage calibrator and Model 5156 Electrometer Calibration standard to the
Model 6514 INPUT jack, as shown in Figure 19-3. Initially, make connections to the
100GΩ resistance.
8. Press ENTER. The instrument will prompt a full-scale calibration current:
20.00000 PA
9. Set the calibrator voltage to +2.000000V. Calculate the actual calibration current from
the calibrator voltage and the actual standard resistor value: I = V/R. Adjust Model 6514
display to agree with the calculated current. Allow the settling time listed in Table 19-5,
then press ENTER.
10. Model 6514 will prompt for the negative full-scale calibration point:
-20PA CAL
11. Press ENTER. The instrument will prompt for the negative full-scale current:
-20.00000 PA
12. Set the calibrator output voltage to -2.000000V, then calculate the calibration current
from the calibrator voltage and standard resistor value: I = V/R. Adjust Model 6514 dis-
play to agree with the calculated current. Allow the settling time listed in Table 19-5,
then press ENTER to complete calibration of the present range.
13. Press EXIT to return to normal display.
14. Repeat steps 1 through 13 for the 200pA through 2µA ranges using Table 19-5 as a
guide. Be sure to make connections to the correct standard resistor.