3. Send the appropriate commands to select the function and range to be calibrated. For ex-
ample, the following commands select the volts function and the 2V range:
:SENS:FUNC ‘VOLT’
:SENS:VOLT:RANG 2
4. Make appropriate connections, then send the commands for each calibration point for
the selected function and range. For example, send the following commands for the 2V
range:
:CAL:PROT:SENS 0
:CAL:PROT:SENS 2
:CAL:PROT:SENS –2
NOTE Be sure the appropriate calibration signal is applied to the INPUT jack before send-
ing the command for each calibration point. See Section 19 for details.
5. Repeat steps 3 and 4 for each function and range.
6. After all functions and ranges are calibrated, send the commands to program the calibra-
tion dates; for example:
:CAL:PROT:DATE 1998,12,15
:CAL:PROT:NDUE 1999,12,15
7. Finally, send the following commands to save calibration constants and then lock out
calibration:
:CAL:PROT:SAVE
:CAL:PROT:LOCK
Calibration Options H-5