Range, Units, Digits, Rate, and Filters 6-3
Every time an autorange occurs, a search for every available range of the selected function is
performed. The time it takes to perform the search could slow down range change speed signif-
icantly. For V, I and ( measurements, upper and/or lower autorange limits can be set to reduce
search time. For Q measurements, the instrument will only autorange between the two higher
charge ranges (high range group), or between the two lower charge ranges (low range group).
NOTE Range limits and groups are not in effect for manual ranging. Every range is accessi-
ble with manual range selection.
Autorange limits for V, I and
Search time for V, I and can be reduced by setting upper and/or lower autorange limits. For
example, if you know the maximum input will be around 1µA, you can set the upper current
range limit to 2µA. This eliminates the 20µA, 200µA, 2mA and 20mA ranges from the search,
therefore, increasing the range change speed. Should the input exceed 2.1µA, the “OVER-
FLOW” message will be displayed.
Perform the following steps to set upper and/or lower autorange limits.
1. Select the V, I or function.
2. Press SHIFT and then one of the following RANGE keys:
a. Press the RANGE
key to display the present UPPER range limit.
b. Press the RANGE
key to display the present LOWER range limit.
3. Use the RANGE
and keys to display the desired limit.
4. Press ENTER.
NOTE If you attempt to select an incompatible range limit, it will be ignored and “TOO
LARGE” or “TOO SMALL” will be displayed briefly. For example, if the lower range
limit is 20V, trying to set the upper limit to 2V will result in the “TOO SMALL” error.
Autorange groups for Q
To optimize range change speed for charge measurements, the instrument will only autorange
between two ranges. With the high range group selected, the instrument can only autorange
between the 2µC and 20µC ranges. With the low range group selected, the instrument can only
autorange between the 20nC and 200nC ranges.
If the HIGH range group is presently selected and the instrument is on the 20nC or 200nC
range (autorange disabled), pressing the AUTO range key will initially select the 2µA range and
then enable autorange. If the range group is then changed to LOW, the instrument will initially
go to the 200nC range with autorange enabled. Therefore, the instrument will always stay within
the selected range group with autorange enabled.
NOTE With the low range group selected, the “OVERFLOW” message will be displayed
when the input signal exceeds 210nC.
Perform the following steps to select autorange group for Q:
1. Select the Q function.
2. Press Shift and then one of the follow RANGE keys:
a. Press the RANGE
key to select the HIGH range group (2µC and 20µC).
b. Press the RANGE
key to select the LOW range group (20nC and 200nC).

Table of Contents

Other manuals for Keithley 6514

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6514 and is the answer not in the manual?

Keithley 6514 Specifications

General IconGeneral
BrandKeithley
Model6514
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals