9.1.3 Weighing with tare
Taring
The dead weight of any weighing container may be tared away by pressing a button,
so that the following weighing procedures show the net weight of the goods to be
weighed.
Place the weighing container on the weighing pan and close the draft shields,
as required.
Wait until the stability display appears ( ), then press . Zero display
and indicator (Net) appear. The weight of the container is now internally
saved.
Weigh the material.
Wait until the stability display appears ( ).
Read net weight.
When the balance is unloaded the saved taring value is displayed with
negative sign.
Taring negative values is not permitted. Any attempt to tare negative
values will result in error message “Err 3“
The taring process can be repeated any number of times. The limit is
reached when the whole weighing range is exhausted.
You can assign the t
are weight to a product in the database. Selecting
the product will automatically load the tare weight, see chap. 15.2.1.
Automatic tare function
When the automatic tare function is enabled <P2.2.2 AUTOTARA YES > the weight
on the pan will first be saved as tare, see chap. 9.3.6