EasyManuals Logo
Home>KERN>Scales>AEJ-C Series

KERN AEJ-C Series User Manual

KERN AEJ-C Series
Go to English
513 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #331 background imageLoading...
Page #331 background image
AEJ-C/AES-C/PLJ-C-BA-e-1621 159
29.6 Detailed information to communication logs
29.6.1 Zeroing
Command: Z CR LF
Possible replies:
Z˽A CR LF
Z˽D CR LF
Command accepted, command executing
Command execution completed
Z
˽
A CR LF
Z˽^ CR LF
Command accepted, command executing
Command valid but zero reset range was exceeded
Z˽A CR LF
Z˽E CR LF
Command accepted, command executing
Time limit for stabilisation of weighing balance display was
exceeded
Z˽ CR LF
Command valid but cannot be executed at the moment
29.6.2 Taring
Command: T CR LF
Possible replies:
T˽A CR LF
T˽D CR LF
Command accepted, command executing
Command execution completed
T
˽
A CR LF
T˽^ CR LF
Command accepted, command executing
Command valid but tare range was exceeded
T˽A CR LF
T˽E CR LF
Command accepted, command executing
Time limit for stabilisation of weighing balance display was
exceeded
T˽I CR LF
Command valid but cannot be executed at the moment
29.6.3 Query tare value
Command: OT CR LF
Response:
1 2 3 4 - 12 13 14 - 16 17 18 19
O T
˽
Tare
˽
Unit
˽
CR LF
Tare value: - 9 characters right aligned

Table of Contents