AEJ-C/AES-C/PLJ-C-BA-e-1621 159
29.6 Detailed information to communication logs
29.6.1 Zeroing
Command: Z CR LF
Possible replies:
Z˽A CR LF
Z˽D CR LF
Command accepted, command executing
Command execution completed
Z
˽
A CR LF
Z˽^ CR LF
Command accepted, command executing
Command valid but zero reset range was exceeded
Z˽A CR LF
Z˽E CR LF
Command accepted, command executing
Time limit for stabilisation of weighing balance display was
exceeded
Z˽ CR LF
Command valid but cannot be executed at the moment
29.6.2 Taring
Command: T CR LF
Possible replies:
T˽A CR LF
T˽D CR LF
Command accepted, command executing
Command execution completed
T
˽
A CR LF
T˽^ CR LF
Command accepted, command executing
Command valid but tare range was exceeded
T˽A CR LF
T˽E CR LF
Command accepted, command executing
Time limit for stabilisation of weighing balance display was
exceeded
T˽I CR LF
Command valid but cannot be executed at the moment
29.6.3 Query tare value
Command: OT CR LF
Response:
1 2 3 4 - 12 13 14 - 16 17 18 19
O T
˽
Tare
˽
Unit
˽
CR LF
Tare value: - 9 characters right aligned