160 AEJ-C/AES-C/PLJ-C-BA-e-1621
Unit: - 3 characters left aligned
The tare value will always be shown in the adjustment unit.
29.6.4 Set tare value
Command: UT˽TARA CR LF, (TARA = tare value)
Possible replies:
UT˽OK CR LF
Command executed
UT˽I CR LF
Command valid but cannot be executed at the moment
ES CR LF
Invalid entry
• Mark decimal places with a point
29.6.5 Send stable weighing value in standard weighing unit
Command: S CR LF
Possible replies:
S˽A CR LF
S˽E CR LF
Command accepted, command executing
Time limit for stabilisation of weighing balance display was exceeded
S˽I CR LF
Command valid but cannot be executed at the moment
S
˽
˽
Signs Weight
˽
Unit CR LF