EasyManuals Logo

KERN AEJ-C Series User Manual

KERN AEJ-C Series
Go to English
513 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #332 background imageLoading...
Page #332 background image
160 AEJ-C/AES-C/PLJ-C-BA-e-1621
Unit: - 3 characters left aligned
The tare value will always be shown in the adjustment unit.
29.6.4 Set tare value
Command: UT˽TARA CR LF, (TARA = tare value)
Possible replies:
UT˽OK CR LF
Command executed
UT˽I CR LF
Command valid but cannot be executed at the moment
ES CR LF
Invalid entry
Setting tare value:
Mark decimal places with a point
Without weighing unit
29.6.5 Send stable weighing value in standard weighing unit
Command: S CR LF
Possible replies:
S˽A CR LF
S˽E CR LF
Command accepted, command executing
Time limit for stabilisation of weighing balance display was exceeded
S˽I CR LF
Command valid but cannot be executed at the moment
1
2-3
4
5
6
7-15
16
17-19
20
21
S
˽
Stability
sign*
˽
Signs Weight
˽
Unit CR LF