29.6.19 Enter reference weight (percentage weighing)
Command: RM˽XXXXX CR LF
Possible replies:
RM˽OK CR LF
Command executed
RM˽I CR LF
Command valid but cannot be executed at the moment (e. g.
incorrect operating mode)
ES CR LF
Syntax error (incorrect weight format)
29.6.20 Confirm / simulation “Press PRINT-key
Command: SS CR LF
Command SS CR LF is used to save the weighing result automatically to the
database and if an optional printer is connected to print out the result.
29.6.21 Start internal adjustment
Command: IC CR LF
Possible replies:
IC ˽A CR LF
Adjustment running
IC ˽D CR LF
Adjustment finished
IC ˽E CR LF
Command valid but cannot be finished at the moment (e. g .
range or time limit for stability display exceeded)
IC ˽I CR LF
Command valid but cannot be executed at the moment
29.6.22 Disable automatic internal adjustment
Command: IC1 CR LF
Possible replies:
IC1 ˽OK CR LF
Command executed
IC1 ˽I CR LF
Command valid but cannot be executed at the moment (Not
available in weighing balances with type approval)
IC1 ˽E CR LF
Command valid but not available at the moment