AEJ-C/AES-C/PLJ-C-BA-e-1621 113
The weighing balance will now be ready for animal weighing:
Select application < ANIMAL WEIGHING >, see chap.16.1.
When using a weighing container tare with TARE-key.
Place load on pan and start the measuring cycle with the function key
assigned to the <START> function (display example F1)
Wait until the result is displayed. Remove load and press the UNITS-key. The
weighing log will be printed automatically if an optional printer is connected
(printout examples see chap. 14.2.2).
Then the balance is ready for further measurements.
21.2 Animal weighing with automatic start
Previous adjustments
Menu item Adjustment Description
<P2.9.7 AUTOSTART> <YES> The measuring cycle will start
automatically as soon as the load is placed
on the pan and the weight exceeds the set
threshold (minimum weight).
<P2.9.7 MEAN. Z.> selectable
1 – 90 s
Duration after which the weighing balance
calculates the mean value of several
weighing results.
<P2.9.7 THRESHOLD> User-defined Definition of minimum weight that starts the
measuring cycle automatically when
exceeded.