AEJ-C/AES-C/PLJ-C-BA-e-1621 63
13.2.2 <P1.2 EXT. JUST.> (AES models)
Carry out adjustment with the help of the recommended adjustment weight
(see chapter 1 “Technical Specifications”).
Observe stable environmental conditions.
A warming up time (see chapter 1)
is required for stabilization.
Invoke menu item < P1.2 EXT. JUST. >
Ensure that there are no objects on the sample dish. Press the PRINT-key
and wait until the weighing value for the required adjustment weight is shown.
Place the required adjustment weight carefully in the centre of the sample dish
and press the PRINT key. Wait until the “MASS REMOVAL” is shown.
Take away adjustment weight. If the adjustment was successful, this will be
followed by automatic return to the menu.