94 AEJ-C/AES-C/PLJ-C-BA-e-1621
17.2 Carry out parts counting
Preparation
For assigning two function keys of your choice with the functions <ENTER SAMPLE>
and <DETERMINE SAMPLE> see chap. 17.1 or chap. 9.3.10.
17.2.1 Determination of the reference piece weight by weighing
Select application <PARTS COUNTING>, see chap.16.1. The reference single
weight last used will be shown in the informational line of text. At the first start-
up as value 0.0000g.
Press function key assigned to < DETERMINE SAMPLE > function, (Display
example F1)
Use the navigation keys to select the desired reference number of parts,
selectable 10, 20, 50 or user-defined.
Selecting reference number of parts 10, 20, 50 or 100:
Confirm selection (such as 100 items) by pressing the PRINT-key.
If necessary, tare the empty balance container.