AEJ-C/AES-C/PLJ-C-BA-e-1621 167
29.6.23 Enable automatic internal adjustment
Command: IC0 CR LF
Possible replies:
IC0 ˽OK CR LF
Command executed
IC0 ˽I CR LF
Command valid but cannot be executed at the moment (Not
available in weighing balances with type approval)
29.6.24 Enter serial number
Command: NB CR LF
Possible replies:
NB˽A˽”x“ CR LF
Command accepted, stating serial number
NB˽I CR LF
Command valid but cannot be executed at the moment
x“: Serial number of device
29.6.25 Disable weighing balance keyboard
Command: K1 CR LF
Possible replies:
K1˽I CR LF
Command valid but cannot be executed at the moment
K1˽OK CR LF
Command accepted, command executing
29.6.26 Enable weighing balance keyboard
Command: K0 CR LF
Possible replies:
K0˽I CR LF
Command valid but cannot be executed at the moment
K0˽OK CR LF
Command accepted, command executing