AEJ-C/AES-C/PLJ-C-BA-e-1621 39
9.3 Settings <P2.2 WEIGHING>
Navigation in menu see chap. 8.2
Menu structure:
Menu item
Parameters
Available settings / explanation
P2.2.1
READING
P2.2.1.1
Filter
EXTRA FAST
Weighing balance reacts very sensitively and very fast,
very stable place of installation.
FAST
Weighing balance reacts sensitively and fast,
stable place of installation.
MEDIUM
Standard normal ambient conditions
SLOW
Weighing balance reacts insensitively but slowly, unstable
place of installation.
VERY SLOW
Weighing balance reacts very insensitively, but very slowly,
very unstable place of installation.
P2.2.1.2
Weighing speed
PRECISE
Rest position control precise, unstable place of installation
FAST
Rest position control fast, very stable place of installation
FAST + PRECISE
Rest position fast + precise, stable place of
installation
P2.2.1.3
Autozero
Selectable yes / no
P2.2.1.4
Last digit
ALWAYS
Last decimal place will be shown
NEVER
Last decimal place will be hidden
WHEN STABLE
Last decimal place will only be shown for stable
weighing values
P2.2.1.5
Environment
STABLE
UNSTABLE