56 AEJ-C/AES-C/PLJ-C-BA-e-1621
P6.9
DATE FORMAT
(on logs)
Options:
YYYY.MM.DD / YYYY.DD.MM / DD.MM.YYYY / MM. DD.YYYY
P6.10
TIME FORMAT
(on logs)
Options: 12H / 24H
The selected format AM or PM will be printed on the log next to
the time.
P6.11
GLP AUTO TEST
Repeatability test activated by placing the internal weight
Sequence of operations:
To start the test, confirm the function <P6.9
AUTOTEST GLP> by pressing .
The internal weight will be applied twice, followed by
10 repeat measurements.
Adjustment
Calculation of standard deviation
Automatic log printout
Printout example KERN YKB-01N
------------------- GLP Autotest Rep.-------------------
-------------------------------------------------------------------
-------------------------------------------------------------------
Additional log output by pressing .
You can cancel the test by pressing .
Press back to weighing mode several times.