Page #232 background imageLoading...
Page #232 background image
60 AEJ-C/AES-C/PLJ-C-BA-e-1621
Automatic adjustment is carried out:
When balance was disconnected from the mains.
When the temperature is changing
The moment a change in temperature of 3°C is detected by the temperature
sensor, the weighing scale will carry out a fully automatic adjustment. The
message „CAL“ in combination with a thermometer icon indicates that an
adjustment is about to take place.
at the end of a defined time interval
After a time interval set in the menu <P1.5 AU.TIME.JUST> has lapsed
(selectable 1 12 h) the weighing balance will announce that adjustment is
about to take place by sending message „CAL“ in combination with a clock icon:
In weighing scales with type approval certificate
Menu item <P1.5 AU.TIME.JUST> disabled
Adjustment takes place every 3 hours