Parameter Typical return
Configure AC current measurements, and turn autorange on:
CONF:CURR:AC
CURR:AC:RANG:AUTOON
Autorange thresholds:
Down range at: <10% of range
Up range at:>120% of range.
Selecting a fixed range (see [SENSe:]CURRent:{AC|DC}:RANGe command) disables autoranging.
This parameter is set to its default value after a Factory Reset (*RST command).
[SENSe:]CURRent:{AC|DC}:RANGe {<range> | MIN | MAX | DEF} [, (@<ch_list>)]
[SENSe:]CURRent:{AC|DC}:RANGe? [{(@<ch_list>) | MIN | MAX | DEF}]
Selects a fixed range for AC and DC current measurements.
Parameter Typical return
<range>:
AC current: {100 μA | 1 mA | 10 mA | 100 mA | 1 A}
DCcurrent: {1 μA | 10 μA | 100 μA | 1 mA | 10 mA | 100 mA | 1 A}
Default: AUTO
+1.00000000E-01
Configure AC current measurements using the 1 A range:
CONF:CURR:AC
CURR:AC:RANG 1
Selecting a fixed range ([SENSe:]<function>:RANGe) disables autoranging.
If the input signal is greater than can be measured on the specified fixed range, the instrument displays Overload
on front panel and returns "9.9E37" from the remote interface.
Changing the value of this parameter always changes the measurement resolution. The number of PLCs (NPLC)
stays fixed, and this results in a change of resolution in terms of the units of the measurement. For example, if
you have a measurement with 1 μA resolution on the 100 mA range, changing to the 1 A range results in a res-
olution of 10 μA.
This parameter is set to its default value after a Factory Reset (*RST command).
230
Keysight DAQ970A/DAQ973A Programming Guide
2 SCPI Programming