EasyManuals Logo

Keysight Technologies Infiniium MXR404A User Manual

Keysight Technologies Infiniium MXR404A
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #71 background imageLoading...
Page #71 background image
Testing Performance 6
Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide 71
5 Press [Clear Display] on the oscilloscope and wait for the number of averages
display (top right area of screen) to return to 256. Record the oscilloscope's
mean V avg reading in the Vzero-error column of Table 7 of the Performance
Test Record. Use the Channel dialog box to select 1 M impedance and then
repeat step 3 to step 5. Record the reading in Table 8 for channel 1.
NOTE
For all oscilloscope readings in this procedure, use the mean value in the Measurement Results
pane at the bottom of the screen.
If a question mark appears in front of any values in the Results pane, press [Clear Display] on the
oscilloscope, wait for the number of averages to return to 256, and then record the oscilloscope
reading. The question mark indicates that the oscilloscope could not make a reliable
measurement.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies Infiniium MXR404A and is the answer not in the manual?

Keysight Technologies Infiniium MXR404A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniium MXR404A
CategoryTest Equipment
LanguageEnglish

Related product manuals