92 Keysight N774xC Series Multiport Power Meter Programming Guide
4 Measurement Operations & Settings
Response: None
Example: :trig0:inp ign
Affects: All power meters
Command: :TRIGger[n]:INPut:EDGe?
Syntax: :TRIGger[n]:INPut:EDGe?
Description: Returns the trigger edge detection of the BNC TTL input.
Parameters: None
Response: RISing: trigger on the high edge of the BNC TTL input
FALLing: trigger on the low edge of the BNC TTL input
Example: :TRIG:INP:EDG? → RIS
Affects: All power meters
Command: :TRIGger[n][:INPut:EDGe
Syntax: :TRIGger[n][:INPut:EDGe
Description: Sets the trigger edge detection of the BNC TTL input.
Parameters: 0 or RISing: trigger on the high edge of the BNC TTL input
1 or FALLing: trigger on the low edge of the BNC TTL input
Response: None
Example: :TRIGg:INP:EDG RIS
Affects: All power meters
Command: :TRIGger[n]:OFFSet?
Syntax: :TRIGger[n]:OFFSet?
Description: Returns the number of incoming triggers received before data logging begins.
Parameters: None