EasyManuals Logo

Keysight N774 C Series Programming Guide

Keysight N774 C Series
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #93 background imageLoading...
Page #93 background image
Keysight N774xC Series Multiport Power Meter Programming Guide 93
Measurement Operations & Settings 4
Response: An integer value.
Example: :trig1:offs? -> 5
Affects: All power meters
Command: :TRIGger[n]:OFFSet
Syntax: :TRIGger[n]:OFFSet <value>
Description: Sets the number of incoming triggers received before data logging begins.
Parameters: <value> - an integer value. (maximum possible value is 2147483647)
Response: None
Example: :trig1:offs 5
Affects: All power meters
Command: :TRIGger[n]:OUTPut?
Syntax: :TRIGger[n]:OUTPut?
Description: Returns the condition that causes an output trigger.
Parameters: none
Response: DISabled:
AVGover:
MEASure:
Never
When averaging time period finishes
When averaging time period begins
Example: :trig1:outp?
-> DIS
Affects: All power meters
Command: :TRIGger[n]:OUTPut
Syntax: :TRIGger[n]:OUTPut
Description: Specifies when an output trigger is generated and arms the channel.