monitor name to connect.
The heart rate monitor must first be
properly placed on the client as described in the Position the
Heart Rate Strap section of this manual. Failure to do so may
result in omission of your monitor from the list.
Successful
connection results in the display of the message below.
b) If no signal is visible, a message similar to the one shown below
displays. Adjust the heart rate strap or add moisture or gel to
improve conductivity.
c) If there continues to be no signal, troubleshoot per the Connect
the BLE Heart Rate Monitor section of this User Manual.
Any
problems with heart rate signal must be resolved prior to
connecting the VO
2
Hose to the CardioCoach PRO unit.
9. Connect the VO
2
hose. Have the client begin to breathe through the
hose. Then the following message displays.
10. Connect VO
2
Hose. Connect the VO
2
hose to the VO
2
MAX port of the
CardioCoach PRO.
11. Detect Stable Heart Rate and Breathing. Ask the client to relax and
breathe normally through the mask and hose. Once the CardioCoach
PRO is filled with expired air and receiving a steady heart rate signal, it
will be ready to begin testing (in approximately 1-2 minutes). One or
two initial large breaths will help clear the mixing chamber and speed
up this process. The client should not start exercising until the
countdown is complete.