LACO ORIGINAL OPERATING INSTRUCTIONS TITANTEST
5
© 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807
Danger from suction on inlet
flange
In vacuum leak test operating mode, there is strong suction on the inlet
flange of the device. If parts of the body or other objects are in the vicinity
of the inlet flange, they may be sucked in.
Always use the provided inlet screen filter.
Always keep parts of the body or other objects away from the inlet flange.
Warning: Damage from vapors If condensible gases or vapors are measured, there is a risk of pump corro-
sion and damage.
Do not immediately switch off the device after the measurement. Let the
device run in "Background cleanup" mode for at least another 20 minutes.
Only then will the pump oil be free of condensed vapors.
Warning: Overpressure Overpressure may destroy the vacuum pump, the vacuum system and the
gaskets.
Do not create an overpressure with the device.
Only use the device for leak testing.
Warning: Halogens in gas If gas whose molecules contain halogen is measured with the device, the
cathode layer of the ion source may be attacked. This may result in the burn-
ing out of the cathode.
Do not test gas whose molecules contain halogens such as fluorine or
chlorine.