6. Operation
6.1 Switching on the device
1. Make sure the plug is removed of the exhaust line (see Chapter 5.2: "Re-
moval of the plug (M-Series only)", page 17).
2. Connect the necessary accessories or equipment before switching on
the device.
3. Switch on the device with the mains plug.
The TITANTEST™ will start and execute an auto test.
The start-up time for the turbo pump is about 2 to 3 minutes. The "Run-
up" window can be seen on the display. A progress bar on the display
shows the process flow.
In this phase, you can already open the menus "Setup", "Language" and
"Information".
After the run-up, the display shows the window "Standby".
4. Allow the device to warm up for 30 minutes before making exact mea-
surements or executing a calibration.
** Feature not available in TITANTEST™ Maintenance model
TITANTEST™ Wet Series: Malfunction of the pump system with inserted
plug in the exhaust line
Operating mode Connected to device
Sniff mode Sniffer probe on vacuum chamber flange
Vacuum mode Test object
Auto Test** Filling system (if applicable)