LACO ORIGINAL OPERATING INSTRUCTIONS TITANTEST™
32
© 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807
Lower display limit You can apply a lower limit to the display of the leakage rate in measure-
ment mode. The setting is only effective for the operating mode "Vacuum"
and "Auto Test**".
► Select " Setup > Global settings > Display settings > Lower Display
Limit".
Language The following languages are available for the menus:
English (factory setting)
German
French
Spanish
Russian
► Select " Setup > Global settings > Display settings > Language".
or
► Alternatively, press "Language" in the "RUN-UP" window.
6.3.2 Access control
In the "Access control" window, you can set access rights for different control
areas.
Maintenance enabled
Access to CAL function
Change device PIN
Change menu PIN
► Select " Setup > Global settings > Access control" and the desired
menu.
Maintenance enabled Define authorization for
– access to the maintenance menu
– venting the turbo molecular pump during the run-up of the device. You
require this function for exchanging the lubricant reservoir of the TMP
SplitFlow80. For more information, please refer to the separate operat
-
ing instructions.
** Feature not available in TITANTEST™ Maintenance model
Table 8: Available measurement units
Option Value range
(Min. Max.)
Comment
Lower display limit With unit mbar*l/s:
1E-12 mbar*l/s
1E-11 mbar*l/s
1E-10 mbar*l/s
1E-9 mbar*l/s
The display limit is only effective in the operating
modes "Vacuum" and "Auto Test
**".