LACO ORIGINAL OPERATING INSTRUCTIONS TITANTEST™
28
© 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807
6.2.7 Standby
After the run-up, the "Standby" window is displayed.
Fig. 11: "Standby" window
A number of parameters are shown below the name "Standby".
Venting ► Select the entry "manual" for "Venting" in " Setup > Test settings >
Evacuation time & Vent" to create an additional "Vent" menu in the
"Standby" window.
► In order to start venting, select "Vent" in the "Standby" window.
► Alternatively: If the setting "with STOP" is set for "Venting" in " Setup >
Test settings > Evacuation time & Vent", press the "Start/Stop" button to
vent.
If the setting "Disabled" is set for "Venting" in " Setup > Test settings >
Evacuation time & Vent", you cannot vent.
If the word "Vented" is displayed, the venting process is complete and you
can replace the test object.
** Feature not available in TITANTEST™ Maintenance model
Table 5: Parameters in the "Standby" window
Parameters Meaning Comment
Operating mode Operating mode Can switch between Vacuum, Sniff, Auto Test**
Q
BG
Current background signal Appears if the corresponding option was selected in
the menu "Background ready to start".
Mass Gas Can be switched between
4
He,
3
He, H
2