LACO ORIGINAL OPERATING INSTRUCTIONS TITANTEST
64
© 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807
Wipe the housing with a soft damp cloth.
Use only water to moisten. Avoid cleaners that contain alcohol, fat or oil.
8.3 Information sources
8.3.1 List of documents
For additional information on the below items see TITANTEST™ USB drive.
Operating instructions UNO 006 A
Operating instructions SplitFlow 80
TITANTEST™ accessories
8.3.2 Calling up maintenance information
The device provides information for maintenance.
You can find an overview of the stored information in "Maintenance & Ser-
vice".
Select " Setup > Maintenance & Service".
In case the menus described in the following are not visible, maintenance
is not enabled. See "'Setup > Access control > Maintenance enabled".
Maintenance device The operating hours of the device and the operating hours since the last
maintenance work are displayed.
Select " Setup > Global settings > Maintenance & Service > Mainte-
nance device".
In order to count the hours of operation after completion of maintenance
starting with "0", set the device maintenance interval counter to "0" by
pressing the "Device" button.
The specification "Operating hours device" cannot be changed.
Maintenance warning If you wish to allow or disable the display of maintenance messages, pro-
ceed as follows:
1. Select " Setup > Global settings > Maintenance & Service > Mainte-
nance message".
2. In order to allow the display of maintenance messages, select "Yes", oth-
erwise "No".
3. Save your setting.
Burn-in You can carry out a load test which starts an automatic start-stop cycle.
In the menu " Setup > Test settings > Evacuation time & Vent", the option
"with Stop" must be set for "Venting" (see Chapter 6.4.4: "Define evacuation
time & vent", page 45).
1. Select " Setup > Global settings > Maintenance & Service > Burn-in".
2. Make sure the test port is blank-flanged.