LACO ORIGINAL OPERATING INSTRUCTIONS TITANTEST™
65
© 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807
3. Start the load test either with the function "Start without calibration" or
"Start with calibration".
The active burn-in process is reported in the display.
4. To stop the process, press the button "Start/Stop".
Maintenance components The hours of operation since the last maintenance are listed for the compo-
nents fore pump, turbo pump, and ion source.
► Select " Setup > Global settings > Maintenance & Service > Mainte-
nance components".
► In order to reset the interval counter for the respective component after
maintenance, press the desired button, i.e. "Fore pump", "Turbo pump", or
"Ion source".
A new entry is written in the maintenance interval list.
Maintenance interval list The maintenance interval list shows the date/time of the last maintenance
for the individual components. It also lists the operating hours of the com-
ponents since their last maintenance.
► Select " Setup > Global settings > Maintenance & Service > View main-
tenance interval list".
Service The Service menu is intended exclusively for service technicians. Access to it
is protected by a service PIN. The service PIN cannot be changed.
After entering the correct service PIN, the Service menu may be accessed.
► Select " Setup > Global settings > Maintenance & Service > Service".
Table 31: Burn-in process
Option Comment
Start with calibration Start of following sequence:
– Calibration, Stop, Start
–Venting, Start, Stop
–Venting, Start, Stop
–Venting, Start, Stop
–Venting, Start, Stop
– Calibration, Start, Stop
etc.
Start without calibration Start of following sequence:
Venting, Start, Stop, etc.