LACO ORIGINAL OPERATING INSTRUCTIONS TITANTEST™
70
© 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807
3. Loosen the cover on the left and right side using an 8 mm Allen wrench,
see Chapter 8.4.2: "Removing the cover (M-Series only)", page 69.
4. Lift off the cover.
5. Loosen the grounding cable.
6. Perform a visual inspection of the oil level and discoloration of the oil in
the UNO
006A.
7. If the oil level is below the minimum mark, remove the oil drain plug us-
ing a screwdriver, fill the oil to the max. mark and screw in the oil drain
plug again.
8. Should the oil be strongly discolored, change the oil (see below).
9. Connect the grounding cable.
10. Replace the cover and secure it with an 8 mm Allen wrench on the left
and right side.
11. Insert the mains plug again.
8.4.3.3 Change the oil (M-Series only)
► Change the oil annually.
1. Switch the device off at the mains plug.
2. Pull out the mains plug.
3. Loosen the cover on the left and right side using an 8 mm Allen wrench,
see Chapter 8.4.2: "Removing the cover (M-Series only)", page 69.
4. Lift off the cover.
5. Loosen the grounding cable.
6. Remove the oil drain plug with a screwdriver.
7. Keep an oil collection container ready and point the oil drain hose to the
container.
8. Use a 13 mm open-ended wrench and 5 mm Allen wrench to loosen the
lock screw at the oil drain hose and drain the oil.
9. Screw in the oil drain plug slightly, insert the mains plug again, and
allow the pump to run for 5 to 10
s to drain the remaining oil.
10. Disconnect the mains plug again.
11. Use a 13 mm open-ended wrench and 5 mm Allen wrench to replace the
lock screw at the oil drain hose.
12. Remove the oil drain plug again and fill in 0.4 l (13.5 fl oz) of fresh oil us-
ing a funnel.
13. Replace the oil drain plug.
14. Connect the grounding cable.
15. Replace the cover and secure it with an 8 mm Allen wrench on the left
and right side.
Required material: Pump Oil P3, 0.5 l Ordering no. T10067