LACO ORIGINAL OPERATING INSTRUCTIONS TITANTEST
29
© 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807
Mode select You can switch between the available operating modes.
1. Select the menu "Mode select" in the window "Standby".
2. Select "Measure mode".
3. Choose from
Vacuum
–Sniff
Auto Test**
The current operating mode is displayed in the "Standby" window
(see Chapter 6.2.7: "Standby", page 28).
Alternatively, you can also switch the operating mode in " Setup >
Test settings > Mode & Mass" and also switch between the detectable gases
(see Tab. 19: "Mode & Mass", page 43).
Setup Here you can make different settings for the measurement tasks, but also
Global settings, such as changing the display or editing authorizations.
Select the menu " Setup" in the window "Standby".
For information on the setting options, see Chapter 6.2.1: "Menu tree",
page 22or see Chapter 6.3: "Global settings", page 30.
Calibration [CAL] This function is displayed if calibration is enabled in the "Access control"
menu.
Select the menu "Calibration" in the window "Standby".
Use this function to start the calibration routine (see Chapter 6.4.10: "Execute
calibration", page 49).
Utilities This menu allows you to access the functions
"Background cleanup"
"Check internal calibration leak"
Select "Standby > Utilities".
Check internal calibration leak This function starts the check of the internal calibration leak (see
Chapter 6.4.10.3: "Check calibration with internal calibration leak", page 51).
The function is only available in the operating mode "Vacuum with Mass 4"
(see Tab. 19: "Mode & Mass", page 43).
Select "Standby > Utilities > Check internal calibration leak".
Background cleanup With the background cleanup function, you can start a start-stop cycle for
breaking down an increased helium background or for preventing conden-
sates (see Chapter : "Background cleanup", page 45).
Select "Standby > Utilities > Background cleanup".