11 Technical data
The information has been conveyed in accordance with DIN 12876.
11.1 General data
Table 33: Command Touch remote control unit
Specification Value Unit
Display type TFT color graphic display ---
Display size 5.7 Inches
Display resolution 640 x 480 Pixels
Display resolution 0.01 °C
Setting resolution 0.01 °C
Table 34: Shared device data
Specification Value Unit
Installation and use Indoors ---
Maximum height of installation above sea level up to 2,000 m
Humidity
Maximum relative air humidity 80 % at ambient
temperature of 31 °C and up to 40 °C, 50 % with
linear decrease
---
Ambient temperature range 5 – 40 °C
IP protection level IP 21 ---
Mains voltage fluctuations
Up to ±10 % of the nominal voltage
Protection class for electrical equipment DIN EN 61
140 (VDE 0140-1)
1 ---
Class division according to DIN 12 876-1
‑ Class designation III ---
‑ Identification code
FL (suitable for combustible and non-combustible
liquids)
---
Bath thermostat temperature stability*
±0.01
K
Circulation thermostat temperature stability*
±0.05
K
Storage temperature range 5 – 40 °C
Transportation temperature range
‑ Heating bath thermostat -20 – 50 °C
‑ Cooling bath thermostat -20 – 43 °C
‑ Thermal circulation thermostat -20 – 50 °C
‑ Cooling circulation thermostat -20 – 43 °C
* Measured value according to standard DIN 12876-2 (12/2001)
V6
PRO bath thermostats and circulation thermostats130 / 156