4 Before starting up
4.1 Installation
WARNING!
Danger of device falling or overturning
Crushing, impacts
Do not tilt the device.
Position the device on an even, non-slip surface with a
sucient load carrying capacity.
Do not position the device close to table edges.
The following instruction is relevant for heat thermostats:
Hold on under the heat thermostats to lift and carry them.
The following safety notice is relevant for devices with a refrigerating
machine:
WARNING!
Danger of overpressure from high ambient temperature
Injury, escape of refrigerant
Note the permitted ambient temperature.
The following safety notices are relevant for devices containing natural
refrigerant:
WARNING!
Explosive gas mixture resulting from leakage in refrigerant circuit
Fire, explosion
Devices containing less than 150 g of combustible refrig-
erant are not subject to special installation conditions.
However, a minimum volumetric flow of 1 m³ for every 8 g
of refrigerant is recommended.
WARNING!
Collection of refrigerant in an installation space similar to a
trough
Fire, explosion
Installation of the device in a space similar to a trough is not
permitted.
The following safety notice is relevant for bath thermostats:
V6
PRO bath thermostats and circulation thermostats 31 / 156